Once removed from the refrigerator, cartridges can be stored at room temperature (18-30 ☌ or 64-86 ☏) for up to two weeks. Where do I store the i-STAT 1 cartridges? All i-STAT 1 cartridges should be refrigerated at 2-8 ☌ (35-46 ☏). If it is out of this range, please take proper steps to get this analyzer back to this temperature range. If the i-STAT 1 instrument is exposed to extreme heat or cold, will it still produce reliable results? Operating temperature for i-STAT 1 cartridge testing is 16-30 ☌ (61-86 ☏). How much blood is required to run an i-STAT 1 cartridge? 17-95 µL, depending on the cartridge used.
#I stat clew update software
New JAMS application software allows the i-STAT1 to recognize any newly launched cartridge types and to perform any newly launched feature. New CLEW software-delivered twice a year on a CD-ROM re-establishes these standardization values and incorporates refinements to the internal quality monitoring system. These updates are equivalent to manually adjusting calibration on a traditional laboratory analyzer.
#I stat clew update update
Due to the continuous manufacturing process improvements to the i-STAT 1 System, it is necessary to update standardization values from time to time to maintain long-term consistency of performance. Why does the i-STAT 1 have software updates? The i-STAT 1 System is designed to eliminate operator influence on delivered results. If running an external control is part of your standard lab practices, controls are available for purchase from Abaxis. The quality checks detect improper environmental conditions, analyzer function, cartridge filling, cartridge function, and sensor function. When there is a quality check failure, a message is displayed with the cause and corrective action.
#I stat clew update series
A series of automatic quality checks are performed during each test cycle. Only maintenance required for the i-STAT is a software update every 6 months.ĭo I need to run controls for the i-STAT 1? No. The calibration adjusts the offset of the stored calibration curve.ĭoes the i-STAT 1 require a lot of maintenance? No, not at all. The signals produced by the sensors' responses to the calibrant solution are measured. During the first part of the testing cycle, the calibrant solution is released from its foil pack and is positioned over the sensors. Calibration is performed each time a cartridge requiring calibration is used, automatically.
#I stat clew update Activator
What analytes on the i-STAT 1 system are measured and which are calculated? Measured: Hematocrit (Hct), Blood Urea Nitrogen (BUN), Creatinine (Crea), Ionized Calcium (iCa), Glucose (Glu), Chloride (Cl), Sodium (Na), Potassium (K), pH, Partial pressure of carbon dioxide (PCO2), Partial pressure of oxygen (PO2), Lactate, Activated Clotting time with Celite® activator (ACT), Cardiac Troponin I (cTnI) Calculated: Hemoglobin (Hb), Bicarbonate (HCO3), Total carbon dioxide concentration (TCO2), Oxygen saturation (sO2), Anion gap (AnGap), Base Excess (BE)ĭoes the i-STAT 1 require calibration? No. These ranges are available in the user guide provided with the analyzer. What species does the i-STAT 1 have reference ranges for? Cat, dog, and horse. What is the shelf-life of the i-STAT 1 cartridges? The cartridges are manufactured with 5-11 months shelf life, depending on the stability of the analytes in each cartridge.


What type of sample is used? Whole blood. These cartridges are configured with parameters that can give a clear patient's condition depending on the clinical situation. What kinds of tests can be run on the i-STAT 1 instrument? The i-STAT 1 offers a variety of disposable, single-use cartridges (10) including tests for acid/base analysis, blood gases, chemistry, hematology, electrolytes, and some specialty tests like Lactate, ACT, and Cardiac Troponin I. This difference in signal is then converted to the specific concentration units for the analyte using the analyzer's internal processor. The concentration of the analyte is determined by measuring the difference in signal from the calibrant and the sample. The sensors are either potentiometric, measuring changes in the electric potential, amperometric, measuring changes in electric current, or conductometric measuring changes in conductivity. These coated sensors react to small changes caused by chemical reactions in the sample. A group of thin film electrodes located at the top of a cartridge are coated with chemically sensitive films. Measurement is accomplished using electrochemical sensing technologies. What technology does the i-STAT 1 utilize? Electrochemistry.